Dew Point Sensor — IZMERKON TP 200
Available on request
Delivery across Russia
Order phone: +7(800)777-18-50
Description
IZMERKON TP 200 - sensor for measuring dew point. Its operation is based on a capacitive sensitive element that absorbs moisture from the air. When the amount of moisture on the surface of the sensitive element changes, its electrical capacitance changes. The sensor monitors this change, as well as the ambient temperature, and based on this data, a special algorithm accurately calculates the dew point temperature.
The main task of the device is to control the humidity of compressed air after the dryer. Thanks to continuous real-time monitoring, it helps prevent problems such as pipeline corrosion, equipment icing, and failure of pneumatic components. This ensures reliable and stable operation of all production equipment.
Features and Benefits
- Alarm function: the alarm triggers when the measurement value goes beyond the threshold range;
- Fast response time thanks to a capacitive sensor with an improved operating method;
- Display on some models: 3” industrial color LCD with a convenient graphical menu and real-time display function of dew point temperature, ambient temperature, pressure (if pressure measurement function is available), as well as alarm status.
Applications
- Measuring the dew point temperature of compressed air after adsorption, membrane, and refrigerated dryers;
- Measuring residual humidity/dew point temperature of gases such as oxygen, nitrogen, argon, etc.;
- Measuring residual humidity/dew point temperature after dryers in plastic production.
- Specifications
- Overall dimensions drawing
- Applications
Specifications |
|
| Measurement range | -80...+20°C dew point -60...+60°C dew point -20...+50°C dew point |
| Sensor type | Capacitive |
| Temperature sensor | PT100 |
| Measurement medium | Air, argon, oxygen, nitrogen, carbon dioxide |
| Response time (T90) at 4 l/min | 0...+20 °C dew point ≤ 50 seconds 20...0 °C dew point ≤ 35 seconds |
| Dew point temperature measurement accuracy | ± 1 °C dew point in the range +50...-20 °C dew point ± 2 °C dew point in the range -20...-50 °C dew point ± 3 °C dew point in the range -50...-80 °C dew point |
| Repeatability | ± 1 °C |
| Relative humidity | 0…90 % RH |
| Measurement medium temperature | -30...+70 °C |
| Measurement medium temperature accuracy | ± 0.3 °C |
| Maximum operating pressure | 1.6 MPa |
| Ambient temperature | 0...+60 °C |
| Ambient humidity | 0…95 % RH |
| Protection rating | IP65 |
| Power supply and power consumption | 12…32 V (DC), 0.72 W at 24 V |
| Data transmission | Modbus RTU (RS485) |
| Analog output | 4…20 mA, ≤ 600 Ohm, accuracy: 0.1 % |
| Digital interface | Modbus RTU (RS485), data transfer rate 115200 baud |
| Housing material | Polycarbonate, ABS |
| Display | 3” Color LCD |
| Process connection | G 1/2” thread |
The specified overall dimensions are for reference only and may vary depending
on the version. Exact dimensions must be confirmed when ordering.