Piezoresistive pressure sensor for use in a hydrogen environment — Series 23SY-H2
Available on request
Delivery across Russia
Order phone: +7(800)777-18-50
Description
The 23SY-H2 series is characterized by low embrittlement due to the increased nickel content in the stainless steel used.
Thanks to the gold-plated diaphragm, H2 diffusion in the 23SY-H2 series is minimized. The sensors of the pressure transmitter are compensated across the entire temperature range and provide excellent long-term stability. The 23SY-H2 series products represent a reliable solution with a long service life, high accuracy and safety, meeting all hydrogen requirements.
Technical Solutions
- Isolated and hermetically sealed piezoresistive pressure sensor
- Fully welded design without internal seals
- Direct signal path with wide bandwidth
Advantages
- Stainless steel with increased nickel content to reduce embrittlement
- Gold-plated membrane to minimize hydrogen penetration
- Compensation across the entire temperature range
- Excellent long-term stability
- Option: intrinsically safe design of the 23SY-Ei-H2 pressure transmitter for use in explosive environments
- Specifications
- Overall dimensions drawing
- Applications
Specifications |
||
| Pressure | ||
| Accuracy | ≤ ± 0.25 %FS | Linearity (best fit straight line), pressure hysteresis, repeatability |
| Total error (0…50 °C) | ≤ ± 0.5 %FS | Maximum deviation within the specified pressure and temperature range |
| Total error (-10…80 °C) | ≤ ± 0.7 %FS | |
| Compensated temperature range | -10…80°C | Option: Other additional temperature ranges possible within -40…125°C |
| Long-term stability | ≤± 0.2% FS | Per year under standard conditions, annual recalibration recommended |
| Position dependence | ≤ ± 1.5 mbar | Calibrated in vertical mounting position with downward threaded connection |
| Electrical characteristics | ||
| Connection | 2-wire | 3-wire |
| Analog signal | 4…20 mA | 0…10 V |
| Output signal limit | 3.2…22.3 mA | -1.2…11.2 V |
| Supply voltage 23SY-H2 | 8…32 V (DC) | 13…32 V (DC) |
| Supply voltage 23SY-Ei-H2 | 10…30 V (DC) | 15…30 V (DC) |
| Power consumption | < 5 mA | |
| Environmental conditions | |||
| Measured medium temperatures | -40…125 °C | No icing allowed | |
| Ambient temperature 23SY-H2 | -40…85 °C | ||
| Ambient temperature 23SY-Ei-H2 | See operating manual | ||
| Storage temperature | -20…85 °C | ||
| Protection | IP67 | Connector M12 x 1 | For gauge pressure IP54 |
| IP65 | Connector type A, DIN EN 175301-803-A (DIN 43650) |
Included in delivery. For gauge pressure a cable with built-in capillary is used. |
|
| Notes | • The degree of protection applies when the corresponding counterpart is present. • The structural design for versions with relative pressure (with an opening for communication with the atmosphere) is shown in the corresponding drawing. |
||
| Vibration resistance | 10 g, 10…2000 Hz, ±10 mm | IEC 60068-2-6 | |
| Shock resistance | 50 g, 11 ms | IEC 60068-2-27 | |
| Number of cycles (service life) at 20…25 °C |
10 million cycles | ≤ 400 bar | 0…100 %FS |
| 4 million cycles | > 400…600 bar | ||
| 1 million cycles | > 600…900 bar | ||
Series 23SY-H2