Differential Pressure Transmitter-Regulator — E2408DF
Available on request
Delivery across Russia
Order phone: +7(800)777-18-50
Description
The differential pressure transmitter-controller E2408DF is a representative of the new line of multifunctional measuring instruments PluraSens®. The transmitter provides high accuracy and zero-level drift absence when measuring air pressure and non-aggressive gases.
Differential pressure measurement is carried out using a thermoelement operating on the principle of microthermoanemometry. Compared to membrane sensors, this technology provides a wider dynamic range, as well as higher stability of readings during long-term measurements and reproducibility of measurements, especially near zero values.
Connection of the transmitter to external devices is provided via the RS485 interface, operating on the Modbus RTU protocol and allowing direct connection of the transmitter to the Fieldbus network, as well as via analog outputs installed upon customer request.
Features
- Ranges from -50...+50 Pa to -500...+500 Pa
- Two configurable analog outputs (0-10 V or 4-20 mA)
- High measurement reproducibility
- Zero-level stability
- Two relays (normally closed contacts)
- Digital RS485 interface
- Specifications
- Overall dimensions drawing
- Connection diagram
- Applications
|
Technical Specifications |
||
|
Measured Medium |
Clean dry air, free of dust, oil and aerosols |
|
|
Measurement Range |
E2408-D50 |
-50...+50 Pa |
|
|
E2408-D500 |
-500...+500 Pa |
|
Resolution |
0.025% of full scale |
|
|
Measurement Type |
micro-thermoanemometry |
|
|
Gas Flow Through Sensor |
< 0.2 ml / min. / Pa |
|
|
Zero Drift |
± 0.2 Pa, stability < 0.1 Pa / year |
|
|
Measurement Accuracy |
±1.5% of full scale at +23°C, 1000 hPa |
|
|
Temperature Compensation |
< 0.5% of full scale per 10°C |
|
|
Atmospheric Pressure Compensation |
< 0.1 % of full scale per hPa |
|
|
Power Supply |
11...30 VDC or 90...265 VAC |
|
|
Power Consumption |
< 2 W |
|
|
Digital Interface |
RS485 and Modbus RTU protocols |
|
|
Analog Outputs |
2 × 4-20 mA / 0-10 V, selectable |
|
|
Output Scaling |
configurable within measurement range |
|
|
Relays |
2 × SPST relays (normally open contact), 250 VAC / 30 VDC, max 5 A |
|
|
Relay Setpoints |
configurable within measurement range |
|
|
Housing |
Light gray ABS plastic, wall mounting, protection class IP65 |
|
|
Dimensions |
WxHxD 85x82x55 mm |
|
|
Pressure connection ports |
Ø4.0/4.9 mm, for 4/6 mm tubes |
|
|
Application area |
industrial spaces |
|
|
Operating environment |
-20...+70 °C, <99 %RH |
|